claude-skills/scripts/evals/fixtures/mdsw2/plan.md
Claude fa9071f8d6
feat(ra-qm): medical-device-software trigger wording and second behaviour case
The first description fired on 6 of 12 held-out prompts: Claude answered
62304 class questions itself even when the prompt named the standard. The
reworded description names the questions users ask (how much process a
class needs, what to record about open-source libraries) and fired on 9 of
12 fresh held-out prompts with no false fires on five near misses.

The second behaviour case targets the misreadings the skill corrects (A1
external controls, FDA level assessed before risk controls, SOUP
identification at class A, EN 62304 citation status): 11/12 with the
plugin against 8/12 without over two runs.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01L8KYFKZR1JGTQLjV6nadZG
2026-11-11 02:26:04 +00:00

1.2 KiB

FlowSafe enteral feeding pump: software classification proposal

Markets: EU (MDR class IIb) and US (510(k), class II).

Hazard

A software fault in the rate controller can over-deliver feed. Our risk file rates the worst case as serious injury (aspiration) if nothing else intervenes.

Hardware cut-off

The pump has an independent over-delivery cut-off: a separate hall-effect volume sensor on its own circuit that removes motor power when delivered volume exceeds the set volume by 5%. It shares no code or processor with the application software and was verified in design verification test DVT-114 (pass, 30/30). With it in place, the risk file rates the residual risk acceptable.

Questions from the team

  1. Our consultant says IEC 62304 classification must be done before any risk control measures, so we are class C or at least B. Is that right, or can the hardware cut-off count?
  2. If we end up class A, can we skip software testing altogether?
  3. Does the FDA documentation level follow the 62304 class?
  4. We use six open-source libraries. At class A, do we need to list them at all?
  5. EN 62304 is harmonised under the MDR, so applying it gives us presumption of conformity for the software, correct?