Medical device software is classified, built and reviewed under three
separate questions that teams routinely merge: the IEC 62304 safety class
(assessed after risk controls external to the software system), the FDA
documentation level (assessed before any risk controls) and the FD&C Act
524B cyber-device test (connectivity, not risk). The library cited IEC
62304 in several ra-qm skills but had no skill that answered any of them.

Four deterministic stdlib scripts with blocking exit codes:
- software_safety_classifier.py: 62304 4.3 with the A1 external-control
  path, item inheritance and segregation, FDA Basic/Enhanced, 524B
- lifecycle_gap_check.py: every 62304 activity by class, 4.4 legacy, and
  IEC 81001-5-1 activities with --security; unassessed is not MET
- soup_register_check.py: 8.1.2, 5.3.3/5.3.4, 7.1.3, end of support, the
  81001-5-1 risk-transfer category (ISH1:2025), SBOM cross-check
- trace_matrix_check.py: requirement/test/result links, risk controls to
  hazards and verification, release state

Every claim in the four references is labelled as checked against the
primary text, agreed by two secondary sources (the paywalled 62304 class
table) or unverified (EU harmonisation of EN 62304 and EN IEC 81001-5-1
under the MDR). Ships cs-device-software-lead, /cs-medical-device-software
and /cs-soup-review, a trigger case and a behaviour case.

Also corrects two existing ra-qm skills: regulatory-affairs-head marked
integration testing required for class A and called IEC 62304 harmonised
under the MDR; fda-consultant-specialist still named the level of concern
that FDA replaced in June 2023. ra-qm-skills 2.14.12, compliance-os
2.14.11 (vendors fda-consultant-specialist). Counters and harness mirrors
regenerated.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
Claude-Session: https://claude.ai/code/session_01L8KYFKZR1JGTQLjV6nadZG
2026-11-11 02:15:18 +00:00

1.2 KiB

InfuSmart P2 volumetric infusion pump: software overview

Target markets: EU (MDR class IIb) and US (510(k), class II).

Software system "InfuCore" (runs on the pump's main processor)

Item Function Notes
DoseEngine Calculates flow rate from the prescribed dose, drives the motor A wrong rate can overdose the patient; worst case is death
DoseGuard Re-checks every commanded rate against the drug library limits and stops the motor on a mismatch Same executable and processor as DoseEngine
UI Touchscreen, alarms display, theme and language settings Team claims class A ("it only draws screens"); no architecture document yet
LogSync Sends event logs to a nurse tablet app over Bluetooth Low Energy

There is also a USB-C service port used for firmware updates.

SOUP

Name Version in register Notes
FreeRTOS latest No requirements recorded; nobody has read its known-issues list
NimBLE (Bluetooth stack) 1.6.0
LVGL (graphics) 8.3.11

Our plan: classify everything as class B because DoseGuard catches dosing errors, keep UI at class A, and submit Basic documentation to FDA.