synced 2026-11-27 22:48:39 +00:00
Medical device software is classified, built and reviewed under three separate questions that teams routinely merge: the IEC 62304 safety class (assessed after risk controls external to the software system), the FDA documentation level (assessed before any risk controls) and the FD&C Act 524B cyber-device test (connectivity, not risk). The library cited IEC 62304 in several ra-qm skills but had no skill that answered any of them. Four deterministic stdlib scripts with blocking exit codes: - software_safety_classifier.py: 62304 4.3 with the A1 external-control path, item inheritance and segregation, FDA Basic/Enhanced, 524B - lifecycle_gap_check.py: every 62304 activity by class, 4.4 legacy, and IEC 81001-5-1 activities with --security; unassessed is not MET - soup_register_check.py: 8.1.2, 5.3.3/5.3.4, 7.1.3, end of support, the 81001-5-1 risk-transfer category (ISH1:2025), SBOM cross-check - trace_matrix_check.py: requirement/test/result links, risk controls to hazards and verification, release state Every claim in the four references is labelled as checked against the primary text, agreed by two secondary sources (the paywalled 62304 class table) or unverified (EU harmonisation of EN 62304 and EN IEC 81001-5-1 under the MDR). Ships cs-device-software-lead, /cs-medical-device-software and /cs-soup-review, a trigger case and a behaviour case. Also corrects two existing ra-qm skills: regulatory-affairs-head marked integration testing required for class A and called IEC 62304 harmonised under the MDR; fda-consultant-specialist still named the level of concern that FDA replaced in June 2023. ra-qm-skills 2.14.12, compliance-os 2.14.11 (vendors fda-consultant-specialist). Counters and harness mirrors regenerated. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com> Claude-Session: https://claude.ai/code/session_01L8KYFKZR1JGTQLjV6nadZG
1.2 KiB
1.2 KiB
InfuSmart P2 volumetric infusion pump: software overview
Target markets: EU (MDR class IIb) and US (510(k), class II).
Software system "InfuCore" (runs on the pump's main processor)
| Item | Function | Notes |
|---|---|---|
| DoseEngine | Calculates flow rate from the prescribed dose, drives the motor | A wrong rate can overdose the patient; worst case is death |
| DoseGuard | Re-checks every commanded rate against the drug library limits and stops the motor on a mismatch | Same executable and processor as DoseEngine |
| UI | Touchscreen, alarms display, theme and language settings | Team claims class A ("it only draws screens"); no architecture document yet |
| LogSync | Sends event logs to a nurse tablet app over Bluetooth Low Energy |
There is also a USB-C service port used for firmware updates.
SOUP
| Name | Version in register | Notes |
|---|---|---|
| FreeRTOS | latest | No requirements recorded; nobody has read its known-issues list |
| NimBLE (Bluetooth stack) | 1.6.0 | |
| LVGL (graphics) | 8.3.11 |
Our plan: classify everything as class B because DoseGuard catches dosing errors, keep UI at class A, and submit Basic documentation to FDA.